The Haier HYC-1031GD is a 1031 L (36.4 cu ft) glass-door pharmacy refrigerator built for high-volume vaccine, reagent, and biological-product storage at a controlled 2–8°C with ≤±1.5°C uniformity. Twelve adjustable shelves, a triple-paned frame-heated glass door, and a low 39 dB(A) hydrocarbon (R290) refrigeration system make it a central cold-store workhorse for hospitals, pharmacies, and pharmaceutical facilities.
Key Specifications
| Model | HYC-1031GD |
| Cabinet type | Upright, triple-paned glass door |
| Temperature range | 2°C to 8°C |
| Temperature uniformity | ≤±1.5°C |
| Cooling / defrost | Forced air / auto defrost |
| Refrigerant | R290 |
| Capacity | 1031 L / 36.4 cu ft |
| Shelves | 12 adjustable (12+2, 14 total) plastic-coated wire |
| Controller / display | Microprocessor (0.1°C, 24h min/max), LED |
| Sound level | 39 dB(A) |
| Power supply | 115 V / 60 Hz |
| Climate class | N (16–32°C ambient) |
Application Profile
When a pharmacy or central store has to hold a large vaccine and reagent inventory, splitting it across several small refrigerators multiplies monitoring points, door-opening losses, and validation work. The HYC-1031GD consolidates that into 1031 L in one cabinet while still holding the tight ≤±1.5°C uniformity that regulated cold storage requires — so a facility gets both scale and precision. Twelve adjustable plastic-coated wire shelves organize the tall interior for improved air circulation and flexible layout.
The R290 hydrocarbon system with forced-air circulation and auto defrost runs quietly at 39 dB(A) with low energy consumption, and an anti-freezing thermostat protects contents from accidental freezing. The triple-paned tempered glass door with frame heater resists condensation at 32°C/80% RH, side-mounted LED lighting illuminates the full depth, and dual 25 mm probe ports on each side wall accept independent monitoring equipment. A top-mounted unique-key lock secures the load.
Design Architecture
The digital microprocessor controller reads to 0.1°C and gives 24-hour min/max recall, backed by a comprehensive alarm suite — high/low temperature, high ambient, door ajar, power failure, low battery, sensor error, communication failure, condenser heat, and remote alarm — with audible/visual alert and mute. A remote alarm contact is standard, and optional USB, RS-485, and a printer export cabinet temperature and alarm data for compliance. Four swivel casters with two leveling feet handle the large cabinet's placement.
Frequently Asked Questions
How much does the HYC-1031GD hold? 1031 L (36.4 cu ft) on 12 adjustable shelves — enough to consolidate a large vaccine or reagent inventory in one cabinet.
How tight is the temperature control? ≤±1.5°C uniformity across 2–8°C, with 0.1°C resolution and 24-hour min/max recall.
Does the glass door prevent condensation? Yes — a triple-paned tempered glass door with frame heater stays clear even at 32°C, 80% RH ambient.
Can I export temperature data for audits? Yes — optional USB, RS-485, and a printer capture and export temperature and alarm records; a remote alarm contact is standard.
Is a solid-door version available? Yes — the companion HYC-1031FD is the same cabinet with a solid door.
Ordering & Configuration
The HYC-1031GD is a capital purchase: request a quote for pricing, freight, and siting. It ships palletized. Backed by Haier Biomedical's manufacturer warranty — confirm current warranty, certification, and voltage configuration with your order.
Choose the HYC-1031GD when high-volume 2–8°C storage must live in one precisely controlled, glass-door cabinet for a hospital or central pharmacy. Drop to the 461 L HYC-461GD for departmental volumes, or specify the solid-door HYC-1031FD when door visibility isn't needed.