The Z326-100H-A15S is the 15 mL conical sealable insert for the Hermle Z326-100H sealed bucket on the Z327 / Z327-K chassis 4-place swing-out rotor, sold as a 2-pack. Each insert holds 1 × 15 mL skirted-base conical tube (116 mm length); the 4-bucket rotor configuration takes 2 packs (4 inserts) for full symmetric loading, delivering 4 × 15 mL conical tubes per cycle in hermetically sealed swing-out geometry. The "S" suffix denotes sealable, unlike the standard A15 insert (used without bucket lid), the A15S is engineered to work with the Z326-100H-LID for hermetic 15 mL workflows (BSL-2 + clinical and clinical-research samples requiring aerosol containment).
Key Specifications
| SKU | Z326-100H-A15S |
| Type | Tube insert (1 × 15 mL conical, sealable) |
| Pack quantity | 2 inserts per SKU (full 4-bucket configuration takes 2 units = 4 inserts) |
| Per-insert capacity | 1 × 15 mL conical (skirted-base / Universal format) |
| Per-cycle throughput (full configuration) | 4 × 15 mL conical per cycle |
| Tube format | 15 mL conical Falcon-style (17 mm OD, 120 mm length) |
| Bucket lid compatibility | Sealable: supports Z326-100H-LID for hermetic envelope |
| Compatible bucket | Z326-100H sealed bucket |
| Compatible chassis | Z327, Z327-K (refrigerated), and -E voltage variants |
| Max. operating speed (with insert + tube) | Per Z326-100H spec (5,000 rpm at 4,080 × g maximum) |
| Max. radius | 14.6 cm (0.5 cm radius correction with this insert) |
| Temperature at max speed (Z327-K refrigerated) | 2 °C |
| Geometry | Swing-out (0–90°): flat sediment surface |
| List price | $267 (2-pack) |
Compatibility & When to Order This SKU
Insert pairing: Z326-100H sealed bucket only. Order this SKU when:
- BSL-2 + 15 mL conical workflow: patient samples or clinical-research samples requiring aerosol containment during the 15 mL spin step.
- Cold-chain BSL-2 clinical-blood-component prep: when both refrigeration (Z327-K chassis at 2 °C max-speed temperature) AND biosafety containment (sealed bucket envelope) are required for the workflow.
- Hermetic-seal clinical workflows at institutional throughput: 4 sealed 15 mL tubes per cycle.
- Replacement inserts if a current set is damaged or worn.
Application Profile
Sealable insert mode: distinct from the A15. The A15S is engineered to work with the Z326-100H-LID, creating a hermetic envelope around each 15 mL tube. The standard A15 insert (different SKU) does not support the bucket lid and is used in non-sealable mode. Per the brochure: A15 is marked "* only without bucket lid"; A15S has no such restriction. The sealable mode is its design intent.
Skirted-base conical compatibility. The A15S envelope is dimensioned for the 17 mm OD × 116 mm length skirted-base / Universal format, a slightly different geometry than standard Falcon/Corning conical (~120 mm length). Verify tube format at order if using non-standard 15 mL tube suppliers.
4-bucket symmetric loading. All 4 buckets must use matching A15S inserts for symmetric balance.
Design Architecture
Cold-chain BSL-2 + clinical blood-component prep at 15 mL: clinical-research and hospital lab workflows handling potentially infectious patient samples in 15 mL conical format that require both refrigeration and aerosol containment.
Viral and bacterial prep workflows at 15 mL: viral RNA extraction, bacterial cell-pellet prep, infectious-agent culture aliquots in 15 mL working volume requiring biosafety containment during centrifugation.
Cell-culture aliquoting from BSL-2 stock cultures: when cell-line work involves BSL-2 + agents and the 15 mL aliquoting step needs hermetic containment.
Clinical-research recruitment workflows handling infectious disease cohorts. Patient sample workflows for infectious disease research (HIV, TB, hepatitis, respiratory pathogens) that require containment from collection through cryostorage.
Cross-format BSL-2 Z327-K chassis: labs running mixed BSL-2 workflows across multiple tube formats (10 mL Vacutainer, 15 mL conical, 5/7 mL clinical) on the same chassis benefit from stocking the sealable insert variants together.
Ordering & Configuration
4-bucket symmetric loading. All 4 buckets must use matching A15S inserts.
Hermetic-seal engagement. Per Hermle service instructions, the Z326-100H-LID must be fully engaged on each bucket before each spin to maintain the hermetic envelope at 4,080 × g.
Tube cap confirmation. All conical tubes must be securely capped before spinning at 5,000 rpm. Biosafety containment depends on both the rotor's hermetic seal and the individual tube cap integrity.
Z326-100H-A15S ships as 2 inserts. 15 mL skirted-base conical tubes are operator-supplied. The Z326-100H bucket and Z326-100H-LID are sold separately.
Confirm at quote: bucket pairing (Z326-100H + Z326-100H-LID in inventory), chassis pairing (Z327 or Z327-K), tube format (15 mL skirted-base / Universal, 116 mm length), workflow biosafety driver (BSL-2 + workflow confirms the sealable A15S vs the non-sealable A15), pack count (2 packs for full 4-bucket configuration), and lead time.
Warranty: 5-Year Hermle warranty (3 years full + 2 years drivetrain) on the inserts. Hermetic-seal interface is a planned-inspection item for biosafety workflows.
Z326-100H 15 mL insert family (Z327 / Z327-K chassis):
- Z326-100H-A15S (this product): 1 × 15 mL conical, sealable mode (BSL-2 + biosafety)
- Z326-100H-A15: 2 × 15 mL conical Falcon-style, non-sealable mode (higher per-cycle throughput)
- Z326-100H-A15L: 1 × 15 mL long-tube (130 mm length), non-sealable
- Z326-100H-A15R: 15 mL round-bottom variant
Pick the A15S when biosafety containment is required; pick the A15 when standard non-sealed throughput matters more.